R&S ESU Series Operating Manual page 353

Emi test receiver
Table of Contents

Advertisement

R&S ESU
For
(CAL REFL SHORT and CAL REFL OPEN) and its mechanisms, refer to section
"Calibration of Reflection Measurement" on page
To carry out a calibration for transmission measurements the whole test setup is
through-connected (THRU).
CAL TRANS
The CAL TRANS softkey triggers the calibration of the transmission measurement.
It starts a sweep that records a reference trace. This trace is then used to calculate
the difference for the normalized values.
Fig. 4.21 Calibration curve for transmission measurement
During the calibration the following message is displayed:
After the calibration the following message is displayed:
This message will be cleared automatically after approx. 3 seconds.
Remote command:
4.322
information
on
the
CORR:METH TRAN
Tracking Generator – Option R&S FSU-B9
calibration
of
the
4.327.
Operating Manual 1302.6163.12 - 03
Instrument Functions
reflection
measurement