R&S ESU Series Operating Manual page 644

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Subsequent commands have to be synchronized with *WAI, *OPC or *OPC? to the
end of the auto range process which would otherwise be aborted.
Example:
"POW:ACH:PRES:RLEV;*WAI"
'Adapts the reference level to the measured channel power.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:BANDwidth|BWIDth 10 to 99.9PCT
This command defines the percentage of the power with respect to the total power. This value is the
basis for the occupied bandwidth measurement (command: POWer:ACHannel:PRESet OBW).
The command is only available in the frequency domain (span > 0).
Example:
"POW:BWID 95PCT"
Characteristics: *RST value: 99PCT
SCPI: device-specific
Mode:
A-F
[SENSe<1|2>:]POWer:HSPeed ON | OFF
This command switches on or off the high-speed channel/adjacent channel power measurement. The
measurement itself is performed in the time domain on the center frequencies of the individual
channels. The command automatically switches to the time domain and back.
Depending on the selected mobile radio standard, weighting filters with √cos characteristic or very
steep-sided channel filters are used for band limitation.
The command is only available in the frequency domain (span > 0).
If the high-speed power measurement is switched off, the complete channel/adja-
cent channel power measurement is switched off.
Example:
"POW:HSP ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A-F
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SENSe Subsystem
6.205

Hide quick links:

Advertisement

Table of Contents
loading