R&S ESU Series Operating Manual page 674

Emi test receiver
Table of Contents

Advertisement

R&S ESU
6.19 SYSTem Subsystem
This subsystem contains a series of commands for general functions.
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess 0 to 30
This command changes the GPIB address of the unit.
Example:
"SYST:COMM:GPIB:ADDR 18"
Characteristics: *RST value: - (no influence on this parameter, factory default 20)
SCPI: conform
Mode:
R, A
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator LFEOI | EOI
This command changes the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary data transfers (e.g.
trace data) from the control computer to the instrument, the binary code (0AH) used for <LF> might be
included in the binary data block, and therefore should not be interpreted as a terminator in this
particular case. This can be avoided by changing the receive terminator to EOI.
Output of binary data from the instrument to the control computer does not require such a terminator
change.
Example:
"SYST:COMM:GPIB:RTER EOI"
Characteristics: *RST value: -- (no influence on this parameter, factory default LFEOI)
SCPI: device-specific
Mode:
R, A
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<1|2>:ADDRess 0 to 30
This command changes the GPIB address of the device selected as external generator 1 or 2.
If two generators are connected at the same time to IECBUS 2 their addresses
must be different.
The command is only available with option Ext. Generator Control, R&S FSP-B10.
Example:
"SYST:COMM:GPIB:RDEV:GEN1:ADDR 19"
'Changes the GPIB address of generator 1 to 19
Characteristics: *RST value: 28
SCPI: device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SYSTem Subsystem
6.235