R&S ESU Series Operating Manual page 811

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Trigger
Threshold
Trigger
Signal
Fig. 7.48 Block diagram illustrating signal processing by the R&S ESU
Either the root-mean-square power or the peak power is measured, depending on
whether the RMS detector or the PEAK detector is selected. The function always
uses TRACE 1 of the selected screen.
The following parameters are to be set for this measurement:
During the measurement, each pulse is mapped into a pixel of the screen, i.e. any
change of the trace can be detected only at the left-hand edge of the screen. Maxi-
mum measurement speed is as usual achieved with the display switched off.
In the example below, a GSM pulse sequence of 8 pulses is measured with 5 µs
trigger offset, 434 µs measurement time/pulse, video trigger with 50% trigger thresh-
old, and peak detection:
REM ************************************************************************
Public Sub MultiBurstPower()
result$ = Space$(200)
'--------- R&S ESU default setting -------------------------
CALL SetupStatusReg
'Configure status register
status = viWrite(analyzer, "*RST", 4, retCnt)
'Reset instrument
status = viWrite(analyzer, "INST:SEL SAN", 12, retCnt)
'Change into analyzer mode
status = viWrite(analyzer, "INIT:CONT OFF", 13, retCnt)
7.22
Measurement
Time
Trigger
Offset
Analyzer frequency
Resolution bandwidth
Measurement time per single pulse
Trigger source
Trigger threshold
Trigger offset
Type of power measurement (PEAK, MEAN)
Number of pulses to be measured
Remote Control – Programming Examples
More Complex Programming Examples
Measurement
Time
Trigger
Offset
Trigger
Signal
Operating Manual 1302.6163.12 - 03
Measurement
Time
Trigger
Offset
Trigger
Signal
t

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents