R&S ESU Series Operating Manual page 266

Emi test receiver
Table of Contents

Advertisement

R&S ESU
START MEAS
STOP MEAS
EDIT SWEEP
Configures the sweep list.
LIST
SWEEP LIST
Range Start
Range Stop
Filter Type
RBW
VBW
Sweep Time Mode AUTO
Sweep Time
Ref Level
RF Att Mode
RF Attenuator
PRE AMP
Transd Factor
Limit Check
Limit
Abs Limit Start -13 dBm
Abs Limit Stop -13 dBm
Rel Limit Start -50 dBc
Rel Limit Stop -50 dBc
Range Start
Start frequency of the range (relative to Center Frequency = Carrier Frequency)
Remote command:
Range Stop
Stop frequency of the range (relative to Center Frequency = Carrier Frequency)
Remote command:
Filter Type
Normal (3 dB), EMI (6 dB), Channel, RRC, 5-Pole, 5-Pole-Digital
Remote command:
RBW
Resolution filter bandwidth
Remote command:
VBW
Video filter bandwidth; not applicable for Channel and RRC filters.
Remote command:
Operating Manual 1302.6163.12 - 03
TX BANDWIDTH
RRC FILT (ON OFF)
ALPHA BT
RANGE 1
RANGE 2 Ref RANGE 3
-250 MHz
-2.52 MHz
-2.52 MHz 2.52 MHz
NORMAL
NORMAL
30 kHz
30 kHz
10 MHz
10 MHz
AUTO
270 ms
10 ms
-10 dBm
-10 dBm
AUTO
AUTO
15 dB
15 dB
OFF
OFF
<NONE>
<NONE>
ON
ON
-13 dBm
-13 dBm
-13 dBm
-13 dBm
-50 dBc
-50 dBc
ESP:RANG1:STAR 1000000
ESP:RANG1:STOP 10000000
ESP:RANG1:FILT:TYPE RRC
ESP:RANG1:BAND:RES 5000
ESP:RANG1:BAND:VID 5000000
Instrument Functions
RANGE 4
2.52 MHz
250 MHz
NORMAL
30 kHz
10 MHz
AUTO
270 ms
-10 dBm
AUTO
15 dB
OFF
<NONE>
ON
-13 dBm
-13 dBm
-13 dBm
-50 dBc
-50 dBc
Analyzer Mode
RANGE 5
4.235