[SENSe<1|2>:]CORRection:TRANsducer:COMMent <string>
This command defines the comment for the selected transducer factor. Prior to this command, the
command SENS:CORR:TRAN:SEL must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:COMM 'FACTOR FOR ANTENNA'"
Characteristics: *RST value: '' (empty comment)
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:DATA <freq>,<level>..
This command defines the reference values of the transducer factor selected. These values are
entered as a sequence of frequency/level pairs. The frequencies must be sent in ascending order. Prior
to this command, the command SENS:CORR:TRAN:SEL must be sent. The level values are sent as
dimensionless numbers; the unit is specified by means of the command SENS:CORR:TRAN:UNIT.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:TRANsducer:DATA 1MHZ,-30,2MHZ,-40"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer[:STATe] ON | OFF
This command switches the selected transducer factor on or off. Prior to this command, the command
SENS:CORR:TRAN:SEL must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:DELete
This command deletes the selected transducer factor. Prior to this command, the command SENS:
CORR:TRAN:SEL must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:DEL"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value.
6.162
Remote Control – Description of Commands
Operating Manual 1302.6163.12 - 03