Remote command:
SELECT TRACE
The SELECT TRACE softkey selects the trace on which the CP/ACP measurement
is to be performed. Only activated traces can be selected, i.e. traces not set to
BLANK.
Remote command:
Examples
Measurement of adjacent-channel power for a specific standard:
The adjacent-channel power is to be measured for a signal at 800 MHz with 0 dBm
level in line with IS136.
[PRESET]
[FREQ: CENTER: 800
MHz]
[AMPT: 0 dBm]
[MEAS]
[CHAN PWR / ACP]
[CP/ACP STANDARD:
select IS136: ENTER]
[CP/ACP CONFIG]
[NO. OF ADJ CHAN: 2
ENTER]
[ADJUST SETTINGS]
PREV
[ADJUST REF LVL]
4.194
Create a new user standards with
Select the user standard you just created with
Change the offset
Save the user standard again to keep the current settings.
Set the R&S ESU to the default setting.
Set the center frequency to 800 MHz.
Set the reference level to 0 dBm.
Call the menu for the measurement functions.
Select the channel and adjacent-channel power measure-
ment function. The measurement is performed with the
default settings or a previously defined setting. The sub-
menu for setting the desired new configuration is opened.
Select the NADC (IS136) standard.
Call the submenu for configuration of the adjacent-channel
power measurement.
Select two adjacent channels for the measurement, i.e.
the adjacent channel and the alternate channel are mea-
sured.
Set the optimum span, resolution bandwidth (RBW), video
bandwidth (VBW) and detector automatically for the mea-
surement. The absolute channel power and the relative
power of the adjacent channels are displayed on the
screen.
Change to the main menu for channel power measure-
ment.
Set the reference level equal to the channel power mea-
sured.
SAVE AS USER STD
SELECT USER STD
POW:ACH:PRES:RLEV:OFFS 10dB
SENS:POW:TRAC 1
Operating Manual 1302.6163.12 - 03