TRACe<1|2>:IQ[:STATe] ON | OFF
This command switches the I/Q data acquisition on or off.
The I/Q data acquisition is not compatible with other measurement functions.
Therefore all other measurement functions will be switched off as soon as the I/Q
measurement function is switched on. Additionally a trace display is not possible in
this operating mode. Therefore all traces are set to "BLANK". Finally split screen
operation will automatically be stopped.
Example:
TRAC:IQ ON
'Switches on I/Q data acquisition
TRAC:IQ:ONL ON
'Switches the online output of the I/Q data on.
INIT:CONT ON
'Selects continuous measurement data output.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A-T
TRACe<1|2>:IQ:FILTer:FLATness NORMal | WIDE
This command allows measurement with extended usable signal bandwidth for sample rates in the
range of either 10.2 MHz to 20.4 MHz or 20.4 MHz to 40.8 MHz.
NORMal
Normal operation
WIDE
Extanded usable signal bandwidth, reduced selectivity
Sample Range
10.2 MHz < Sample Rate <= 20.4 MHz
20.4 MHz < Sample Rate <= 40.8 MHz
Example:
"TRAC:IQ:FILT:FLAT WIDE"
Characteristics: *RST value: NORM
SCPI: device-specific
Mode:
A
6.258
Remote Control – Description of Commands
Usable Bandwidth
NORMAL
WIDE
0.8 * Sample Rate
0.9 * Sample Rate
0.68 * Sample Rate
0.8 * Sample Rate
TRACe Subsystem
Operating Manual 1302.6163.12 - 03