R&S ESU Series Operating Manual page 456

Emi test receiver
Table of Contents

Advertisement

R&S ESU
CALCulate<1|2>:DELTamarker<1...4>:MINimum:RIGHt
This command positions the specified delta marker to the next higher minimum value to the right of the
current value (i.e. ascending X values). The corresponding delta marker will be activated first, if
necessary.
Example:
"CALC2:DELT:MIN:RIGH"
'Sets delta marker 1 in screen B to the next higher minimum value to the right of
the current value.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:MINimum:LEFT
This command positions the specified delta marker to the next higher minimum value to the left of the
current value (i.e. descending X values). The corresponding delta marker will be activated first, if
necessary.
Example:
"CALC:DELT:MIN:LEFT"
'Sets delta marker 1 in screen A to the next higher minimum to the left of the current
value.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:LINK ON | OFF
This command switches on and off a link of delta marker 1 to marker 1. If the link is switched on and
the x-axis value of the marker 1 is changed, the delta marker 1 will follow on the same x-position.
The suffix at DELTamarker can only be 1 or not present, because the functionality is only available for
marker 1 and delta marker 1.
Example:
"CALC1:DELT1:LINK ON"
Switches the link of marker1/delta marker 1 on.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe] ON | OFF
This command switches the relative measurement to a fixed reference value on or off. Marker 1 will
be activated previously and a peak search will be performed, if necessary. If marker 1 is activated, its
position becomes the reference point for the measurement. The reference point can then be modified
with commands CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:X and to:RPOint:Y
independently of the position of marker 1 and of a trace. It is valid for all delta markers in the selected
measurement window as long as the function is active.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.17