R&S ESU Series Operating Manual page 650

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]SCAN<1 to 10>:TIME 10 µs to 100 s
This command defines the measurement time of the receiver subscan.
With option ESU-K53 (Time Domain Scan) the measurement time range can be set from 10 µs to
100 s and is dependent on the selected resolution bandwidth.
Example:
":SCAN1:TIME 1 ms"
'Sets the measurement time of scan range 1 to 1 ms
Characteristics: *RST value: 1 ms (range 1)
100 µs (range 2)
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:ATTenuation dB
This command defines the RF attenuation for the selected receiver scan range.
Example:
":SCAN1:INP:ATT 30dB"
'Sets the input attenuation of scan range 1 to 30 dB
Characteristics: *RST value: 10 dB
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:ATTenuation:AUTO ON | OFF
This command switches on or off the autoranging function in the selected receiver scan range.
Example:
":SCAN1:INP:ATT:AUTO ON"
'Activates the auto ranging function for scan range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:GAIN[:STATE] ON | OFF
This command switches on or off the preamplifier in the selected receiver scan range.
Example:
":SCAN1:INP:GAIN:STAT ON"
'Activates the preamplifier for scan range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
to dB
min
max
SENSe Subsystem
6.211

Hide quick links: