6.6 CALibration Subsystem
The commands of the CALibration subsystem determine the data for system error correction in the
instrument.
CALibration[:ALL]?
This command initiates the acquisition of system error correction data. A "0" is returned if the
acquisition was successful.
During the acquisition of correction data, the instrument does not accept any
remote control commands, except
*RST
CALibration:ABORt
In order to recognize when the acquisition of correction data is completed, the MAV bit in the status
byte can be used. If the associated bit is set in the Service Request Enable Register, the instrument
generates a service request after the acquisition of correction data has been completed.
Example:
"*CLS"
'Resets the status management.
"*SRE 16"
'Enables MAV bit in the Service Request Enable Register.
"*CAL?"
'Starts the correction data recording and then a service request is generated.
Characteristics: *RST value: -
SCPI: conform
Mode:
R, A
CALibration:ABORt
This command aborts the acquisition of correction data and restores the last complete correction data
set.
Example:
"CAL:ABOR"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
6.98
Remote Control – Description of Commands
CALibration Subsystem
Operating Manual 1302.6163.12 - 03