R&S ESU Series Operating Manual page 600

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]CORRection:TRANsducer:GENerate <name>
This command generates a transducer factor <name> using normalized trace data. The function is
only available when normalization is switched on.
Parameter:
<name>::= Name of the transducer factors as string data with up to 8 characters.
Example:
"CORR:TRAN:GEN 'FACTOR1'"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]CORRection:TRANsducer:SELect <name>
This command selects the transducer factor designated by <name>. If <name> does not exist yet, a
new transducer factor is created.
This command must be sent prior to the subsequent commands for modifying/acti-
vating transducer factors.
Parameter:
<name>::= Name of the transducer factor in string data form with a maximum of 8
characters.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:UNIT <string>
This command specifies the unit for the selected transducer factor. Prior to this command, the
command SENS:CORR:TRAN:SEL must be sent.
Parameter:
<string>::= 'DB' | 'DBM' | 'DBMV' | 'DBUV' | 'DBUV/M' | 'DBUA' | 'DBUA/M' | 'DBPW'
| 'DBPT'
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:UNIT 'DBUV'"
Characteristics: *RST value: 'DB'
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:SCALing LINear | LOGarithmic
This command defines whether the frequency scaling of the transducer factor is linear or logarithmic.
Prior to this command, the command SENS:CORR:TRAN:SEL must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:SCAL LOG"
Characteristics: *RST value: LINear
SCPI: device-specific
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SENSe Subsystem
6.161