R&S ESU
SIGNAL
The SIGNAL STATISTIC softkey opens a submenu for measurement of signal
STATISTIC
statistics.
APD ON | OFF
CCDF ON | OFF
PERCENT MARKER
RES BW
NO OF SAMPLES
SCALING
ADJUST SETTINGS
CONT MEAS
SINGLE MEAS
Side menu
GATED TRIGGER
GATE RANGES
In the submenu measurement of amplitude probability density (APD) and comple-
mentary cumulative distribution (CCDF) can be selected alternately. Only one of the
signal statistic functions can be switched on at a time.
In default mode all statistic functions are switched off.
With a statistic function switched on the R&S ESU is set into zero span mode auto-
matically.
The R&S ESU measures the statistics of the signal applied to the RF input with the
resolution bandwidth set. In order not to influence the peak amplitudes the video
bandwidth is automatically set to 10 times the resolution bandwidth. The sample
detector is used for detecting the video voltage.
APD ON | OFF
The APD ON/OFF softkey switches on or off the amplitude probability distribution
function. When the APD function is switched on, the CCDF function is switched off
automatically.
Remote command:
CCDF ON | OFF
The CCDF ON/OFF softkey switches on or off the complementary cumulative distri-
bution function. When the CCDF function is switched on, the APD function is
switched off automatically.
Remote command:
4.202
!
X-AXIS REF LEVEL
X-AXIS RANGE
Y-UNIT %/ABS
Y-AXIS MAX VALUE
Y-AXIS MIN VALUE
ADJUST SETTINGS
DEFAULT SETTINGS
MEAN PWR POSITION
CALC:STAT:APD ON
CALC:STAT:CCDF ON
Instrument Functions
Analyzer Mode
Operating Manual 1302.6163.12 - 03