R&S ESU Series Operating Manual page 655

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]SWEep:EGATe:LENGth 0 to 100 s
In case of edge triggering, this command determines the time interval during which the instrument
sweeps.
Example:
"SWE:EGAT:LENG 10ms"
Characteristics: *RST value: 0s
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:EGATe:SOURce EXTernal | IFPower
This command toggles between external gate signal and IF power signal as a signal source for the
gate mode. If an IF power signal is used, the gate is opened as soon as a signal at > -20 dBm is
detected within the IF path bandwidth.
Example:
"SWE:EGAT:SOUR IFP"
'Switches the gate source to IF power.
Characteristics: *RST value: IFPower
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:EGATe:TRACe<1..3>COMMent
To enter a comment string in the gate range table.
The suffix after TRACe indicates the trace.
Example:
"SWE:EGAT:TRAC2:COMMent"
'To enter a comment in trace 2.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:EGATe:TRACe<1..3>STATe<1..3> ON | OFF
The gate range in the gate range table is set on or off.
The suffix after TRACe indicates the trace.
The suffix after STATe indicates the range.
Example:
"SWE:EGAT:TRAC2:STAT1 ON"
'Range 1 for trace 2 is switched on.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
6.216