CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command is used to query the result of the measurement of the positive peak value with active
peak hold function. The query is only possible if the peak hold function has been activated previously
using CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PHOLd.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:PHOL ON"
'Switches on the measurement of the peak value in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:PHOL:RES?"
'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS[:STATe] ON | OFF
This command switches on or off the measurement of the effective (RMS) power in the selected
measurement window. If necessary the function is switched on previously.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
Example:
"CALC2:MARK:FUNC:SUM:RMS ON"
'Switches on the function in screen B.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS:RESult?
This command queries the result of the measurement of the RMS power value in the selected
measurement window.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.77