6.13 INSTrument Subsystem
The INSTrument subsystem selects the operating mode of the unit either via text parameters or fixed
numbers. Only operating modes available for the base unit and the models and the options described in
this manual (see
"Documentation Overview" on page
modes refer to the corresponding separate manuals.
INSTrument[:SELect] SANalyzer | RECeiver | IFANalyzer | ADEMod | APD
This command enables you to switch between modes by entering the mode designation.
Parameter:
SANalyzer: spectrum analyzer
RECeiver: Receiver mode
IFANalyzer: IF analysis mode
APD: Amplitude Probability Distribution
Example:
"INST SAN"
'Switches the instrument to SPECTRUM.
Characteristics: *RST value: RECeiver
SCPI: conform
Mode:
R, A
A changeover to a mode other than SANalyzer and RECeiver is only possible if
the corresponding option is installed.
INSTrument:NSELect <numeric value>
This command enables you to switch between the modes by using numbers.
Parameter:
1: Analyzer mode
3: FM demodulator mode
6: Receiver mode
23: IF Analyzer
24: APD Mode
Example:
"INST:NSEL 1"
'Switches to the analyzer mode.
Characteristics: *RST value: 6
SCPI: conform
Mode:
R, A
A changeover to a mode is only possible if the corresponding option is installed.