6Selection Of Detectors For Final Measurement - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
GET MAXHOLD
The GET MAXHOLD softkey accepts the highest level measured during the HOLD
SCAN state as a result for the final measurement and continues the prescan. (The
level value in question is displayed as a small bar in the bargraph.)
Remote command:
MEASURE
The MEASURE softkey continues with the final measurement. The final measure-
ment starts at the next frequency entry on the peak list or at a marked frequency if
one or several lines have been skipped with the SKIP FREQUENCY softkey.
Remote command:
STOP
The STOP FINAL MEAS softkey halts the final measurement. The final measure-
FINAL MEAS
ment starts from the beginning on a new start. The previously collected data is lost.
Remote command:
The peak list available after the final measurement:
4.5.4.6
Selection of Detectors for Final Measurement
The selection of detectors for the final measurement is made in the right side menu
TRACE DETECTOR (see section
on page
The detectors to be used for the final measurement can be set here for each trace,
i.e. any combination of prescan and final measurement is possible. The required
flexibility is thus obtained for the diverse test specifications which are covered by
means of the R&S ESU.
In the THRESHOLD SCAN mode, with the NB/BB DISCR function active, the
R&S ESU automatically selects the detector to be used in the final measurement on
the basis of the results obtained in the prescan.
4.46
--
--
--
EDIT PEAK LIST (Final Measurement Results)
Trace1:
014QP
Trace2:
014AV
TRACE3:
---
FREQUENCY
TRACE
1
Average
80.0000 MHz
2
Average
89.4800 MHz
1
Quasi Peak
98.5200 MHz
2
Average
98.5200 MHz
1
Quasi Peak
100.7200 MHz
2
Average
102.3200 MHz
1
Quasi Peak
113.2400 MHz
2
Average
116.9200 MHz
1
Quasi Peak
125.8800 MHz
2
Average
125.8800 MHz
1
Quasi Peak
138.4800 MHz
2
Average
138.4800 MHz
2
Average
144.0400 MHz
2
Average
167.0400 MHz
2
Average
176.2400 MHz
1
Quasi Peak
200.4800 MHz
2
Average
200.4800 MHz
1
Quasi Peak
210.2800 MHz
2
Average
226.5600 MHz
2
Average
230.0000 MHz
4.70).
LEVEL dBpT
29.99
35.64
49.94
48.32
55.33
50.86
42.50
44.44
54.91
53.86
41.83
39.38
40.77
44.82
46.56
50.93
48.27
58.71
59.07
46.90
"Selection and Setting of Traces – TRACE Key"
Operating Manual 1302.6163.12 - 03
Instrument Functions
Receiver Mode
DELTA LIMIT dB
-9.25
-4.09
-0.22
8.15
5.07
10.53
-8.26
3.53
3.68
12.64
-9.81
-2.25
-1.04
2.37
3.87
-2.31
5.02
5.25
15.29
3.05