R&S ESU Series Operating Manual page 435

Emi test receiver
Table of Contents

Advertisement

R&S ESU
5.7.3.11
STATus:QUEStionalble:TRANsducer Register
This register indicates that a transducer hold point is attained (bit 15) and what
range is to be swept next (bit 0 to 10). The sweep can be continued with command
"INITiate<1|2>:CONMeas".
It can be queried with commands
tion?"
Bit
No.
0
1
2
3
4
5
6
7
8
9
10 to
13
14
15
5.30
and "STATus:QUEStionable:TRANsducer[:EVENt]?".
Meaning
Range 1
This bit is set when subrange 1 is attained.
Range 2
This bit is set when subrange 2 is attained.
Range 3
This bit is set when subrange 3 is attained.
Range 4
This bit is set when subrange 4 is attained.
Range 5
This bit is set when subrange 5 is attained.
Range 6
This bit is set when subrange 6 is attained.
Range 7
This bit is set when subrange 71 is attained.
Range 8
This bit is set when subrange 8 is attained.
Range 9
This bit is set when subrange 9 is attained.
Range 10
This bit is set when subrange 10 is attained.
not used
Subrange limit
This bit is set when the transducer is at the point of changeover from one range to another.
This bit is always 0.
Remote Control – Basics
Status Reporting System
"STATus:QUEStionable:TRANsducer:CONDi-
Operating Manual 1302.6163.12 - 03