Order of signals detected:
PEAK:
NEXT PEAK:
NEXT PEAK:
or
PEAK:
NEXT PEAK LEFT:
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT: Signal 2 (no further signal detected)
With 6 dB peak excursion, all signals will be detected with NEXT PEAK and NEXT
PEAK RIGHT / NEXT PEAK LEFT.
Order of signals detected:
PEAK:
NEXT PEAK:
NEXT PEAK:
NEXT PEAK:
or
PEAK:
NEXT PEAK LEFT:
NEXT PEAK RIGHT: Signal 1
NEXT PEAK RIGHT: Signal 2
NEXT PEAK RIGHT. Signal 4
AUTO MAX
The AUTO MAX PEAK / AUTO MIN PEAK softkeys add an automatic peak search
PEAK
action for marker 1 at the end of each particular sweep. This function may be used
AUTO MIN
during adjustments of a device under test to keep track of the current peak marker
PEAK
position and level.
The current marker search limit settings (LEFT LIMIT, RIGHT LIMIT, THRESHOLD,
EXCLUDE LO) are taken into account.
Remote command:
4.168
Signal 1
Signal 2
Signal 2 (no further signal detected)
Signal 1
Signal 1 (no further signal detected)
Signal 1
Signal 2
Signal 3
Signal 4
Signal 1
Signal 3
CALC:MARK:MAX:AUTO ON | OFF
CALC:MARK:MIN:AUTO ON | OFF
Operating Manual 1302.6163.12 - 03