6.16.8
SENSe:FREQuency Subsystem
The SENSe:FREQuency subsystem defines the frequency axis of the active display. The frequency axis
can either be defined via the start/stop frequency or via the center frequency and span.
The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
[SENSe<1|2>:]FREQuency:CENTer 0 to f
This command defines the receiver frequency or the center frequency in analyzer mode.
Example:
"FREQ:CENT 100MHz"
Characteristics: *RST value: f
SCPI: conform
Mode:
R, A
[SENSe<1|2>:]FREQuency:CENTer:STEP 0 to f
This command defines the step width of the center frequency.
Example:
"FREQ:CENT:STEP 120MHz"
Characteristics: *RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: conform
Mode:
R, A
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
This command couples the step width of the center frequency to span (span >0) or to the resolution
bandwidth (span = 0) or cancels the couplings.
Parameter:
SPAN = Coupling to frequency display range (for span > 0)
RBW = Coupling to resolution bandwidth (for span = 0)
OFF = manual input, no coupling.
Example:
"FREQ:CENT:STEP:LINK SPAN"
Characteristics: *RST value: SPAN
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor 1 to 100 PCT
This command couples the step width of the center frequency with a factor to the span (span >0) or to
the resolution bandwidth (span = 0).
Example:
"FREQ:CENT:STEP:LINK:FACT 20PCT"
Characteristics: *RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: device-specific
Mode:
A
6.180
max
/2 with f
= maximum frequency
max
max
max
Remote Control – Description of Commands
Operating Manual 1302.6163.12 - 03
SENSe Subsystem