4.5.2
Level Display and RF Input Configuration – AMPT Key
The AMPT key is used to set the input attenuation, the Preamplifier, the auto range
function and the
In addition, the level display range for the scan can be set.
AMPT
RF ATTEN MANUAL
PREAMP ON OFF
LN PREAMP ON OFF
10 DB MIN ON | OFF
AUTO RANGE ON | OFF
AUTOPREAMP ON | OFF
RF INPUT 1 | 2
RF INPUT AC | DC
GRID LEVEL
Side menu
dBmV | dBm | dBmA | dBpW |
dBpT | dBmV | dB*/MHz
*) LN PREAMP ON OFF is only available with option R&S ESU-B24
RF ATTEN
The RF ATTEN MANUAL softkey activates the attenuation entry field.
MANUAL
The attenuation can be set between 0 and 75 dB in 5 dB steps. Other entries are
rounded up to the next higher integer.
To protect the input mixer against inadvertent overload, 0 dB can only be
switched on when the 10 dB MIN softkey is switched off.
Remote command:
Preamplifier
R&S ESU is provided with a switchable preamplifier of 20 dB gain in the frequency
range up to 3.6 GHz. An additional low noise preamplifier which covers the whole
frequency range is available as option R&S ESU-B24.
Switching on the preamplifier reduces the total mark figure of the R&S ESU and thus
improves the sensitivity.
The disadvantage of a poorer large-signal immunity (intermodulation) is reduced by
the connected preselector. The signal level of the subsequent mixer is 20 dB higher
so that the maximum input level is reduced by the gain of the preamplifier. The use
of the preamplifier is recommended when measurements with a maximum sensitiv-
ity are to be performed. On the other hand, if the measurement should be performed
at maximum dynamic range, the preamplifier should be switched off.
4.16
Display
Unit.
GRID RANGE LOG MANUAL
GRID MIN LEVEL
INP:ATT 40 DB
Instrument Functions
Receiver Mode
Operating Manual 1302.6163.12 - 03