R&S ESU Series Operating Manual page 583

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Example:
"MMEM:MSIS 'F:'"
Characteristics: *RST value: "D:'
SCPI: conform
Mode:
R, A
MMEMory:NAME <file_name>
This command defines a destination file for the printout started with the command HCOPy:
IMMediate. In this case the printer output must be routed to destination FILE using the command
"HCOP:DEST 'MMEM' ".
The file name includes indication of the path and may also include the drive name. The file name and
path information comply with DOS conventions.
Parameter:
<file_name> ::= DOS file name
Example:
"HCOP:DEV:LANG BMP"
'Selection of data format.
"HCOP:DEST 'MMEM'"
'Selection of the output device
"MMEM:NAME 'PRINT1.BMP'"
'Selection of file name.
"HCOP:IMM"
'Start of the printout.
Characteristics: *RST value: -
SCPI: conform
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
MMEMory:RDIRectory <directory_name>
This command deletes the indicated directory. The directory name includes indication of the path and
may also include the drive name. The path name complies with DOS conventions.
Parameter:
<directory_name>::= DOS path name
Example:
"MMEM:RDIR 'D:\TEST'"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
MMEMory:SELect[:ITEM]:HCOPy
This command copies the test report settings to the list of partial data records of a device setting to be
stored/loaded.
Example:
"MMEM:SEL:HCOP ON"
Characteristics: *RST value:OFF
SCPI:device-specific
Mode:
R
6.144
Remote Control – Description of Commands
ON | OFF
MMEMory Subsystem