R&S ESU Series Operating Manual page 538

Emi test receiver
Table of Contents

Advertisement

R&S ESU
CALibration:RESult?
This command outputs the results of the correction data acquisition. The lines of the result table (see
chapter
"Instrument
are output as string data separated by commas:
Return value:
"Total Calibration Status: PASSED","Date (dd/mm/yyyy): 12/07/
2006",
"Time: 16:24:54","Runtime:00.06"
Example:
"CAL:RES?"
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R, A
CALibration:STATe ON | OFF
This command determines whether the current calibration data are taken into account by the
instrument (ON) or not (OFF).
Example:
"CAL:STAT OFF"
'Sets up the instrument to ignore the calibration data.
Characteristics: *RST value: -
SCPI: conform
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Functions", section
"Recording the Correction Data – CAL Key" on page
Remote Control – Description of Commands
CALibration Subsystem
4.143)
6.99