If LIM CHECK is set to OFF for all active limit lines, then the limit line check is not
executed and the display window is activated.
Remote command:
Trace - Select the measurement curve to which the limit line is assigned.
The selection of the measurement curve (trace) takes place in an entry window.
Allowed are the integer entries 1, 2 or 3. The default setting is trace 1. If the selected
limit line is not compatible with the assigned measurement curve, then the limit line
is disabled (display and limit check).
Remote command:
NEW LIMIT
See following section
LINE
EDIT LIMIT
See following section
LINE
COPY LIMIT
The COPY LIMIT LINE softkey copies the data file describing the marked limit line
LINE
and saves it under a new name. In this way, a new limit line can be easily generated
by parallel translation or editing of an existing limit line. The name can be arbitrarily
chosen and input via an entry window (max. of 8 characters).
Remote command:
DELETE LIMIT
The DELETE LIMIT LINE softkey erases the selected limit line. Before deletion, a
LINE
message appears requesting confirmation.
Remote command:
X OFFSET
The X OFFSET softkey horizontally shifts a limit line, which has been specified for
relative frequencies or times (x-axis). The softkey opens an entry window, where the
value for shifting may be entered numerically or via the rotary knob.
This softkey does not have any effect on limit lines that represent absolute values
for the x-axis.
Remote command:
Y OFFSET
The Y OFFSET softkey vertically shifts a limit line, which has relative values for the
y-axis (levels or linear units such as volt). The softkey opens an entry window where
the value for shifting may be entered numerically or via the rotary knob.
4.244
CALC:LIM:STAT ON
INIT;*WAI
CALC:LIM:FAIL?
CALC:LIM:TRAC 1
"Entry and Editing of Limit Lines" on page
"Entry and Editing of Limit Lines" on page
CALC:LIM3:COPY 2
or
CALC:LIM3:COPY "GSM2"
CALC:LIM3:DEL
CALC:LIM3:CONT:OFFS 10kHz
Instrument Functions
4.245.
4.245.
Operating Manual 1302.6163.12 - 03
Basic Settings