The permissible setting range is -200 dB to +200 dB in steps of 0.1 dB. Positive off-
sets handle a subsequent amplifier and negative offsets an attenuator pad.
The default setting is 0 dB; offsets <> 0 are marked by the activated enhancement
label LVL.
Remote command:
4.11.2

Transmission Measurement

The transmission characteristic of a two-port network is measured. The external
generator serves as a signal source. It is connected to the input connector of the
DUT. The input of the R&S ESU is fed from the output of the DUT.
Fig. 4.29 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g.
frequency response of connecting cables).
4.11.2.1
Calibration of Transmission Measurement
SOURCE CAL
CAL TRANS
CAL REFL SHORT
CAL REFL OPEN
NORMALIZE
REF VALUE POSITION
REF VALUE
RECALL
SAVE AS TRD FACTOR
The SOURCE CAL softkey opens a submenu comprising the calibration functions
for the transmission and reflection measurement.
For
(CAL REFL SHORT and CAL REFL OPEN) and its mechanisms, refer to section
"Calibration of Reflection Measurement" on page
To carry out a calibration for transmission measurements the whole test setup is
through-connected (THRU).
Operating Manual 1302.6163.12 - 03
SOUR:POW:OFFS -10dB
information
on
the
External Generator Control – Option R&S FSP-B10
calibration
of
the
4.341.