R&S ESU Series Operating Manual page 651

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]SCAN<1 to 10>:INPut:GAIN:AUTO ON | OFF
This command includes the preamplifier in the autoranging function of the selected receiver scan
range.
Example:
":SCAN1:INP:GAIN:AUTO ON"
'Includes the preamplifier in the autoranging function for scan range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:GAIN:LNA[:STATe] ON | OFF
This command switches the low noise preamplifier on or off for a receiver subscan.
(Only with option R&S ESU-B24).
Example:
":SCAN1:INP:GAIN:LNA ON"
'Switches the low noise amplifier on for scan range 1
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:INPut:TYPE INPUT1 | INPUT2
This command defines the input for a receiver subscan.
Example:
":SCAN1:INP:TYPE INPUT2"
'Activates RF input 2 for scan range 1
Characteristics: *RST value: INPUT 1
SCPI: device-specific
Mode:
R
[SENSe<1|2>:]SCAN<1 to 10>:RANGes[:COUNt] 1 to 10
This command determines the number of ranges. With setting 0, the scan is performed using the
current receiver settings and not the settings defined with commands SENSe:SCAN:....
Example:
"SCAN:RANG:COUN 3"
'Sets the number of ranges to 3
Characteristics: *RST value: 0
SCPI: device-specific
Mode:
R
6.212
Remote Control – Description of Commands
Operating Manual 1302.6163.12 - 03
SENSe Subsystem

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents