R&S ESU
Table 4-1
File contents
Type;R&S ESU;
Version;3.90;
Date;02.Aug 2006;
Mode;ANALYZER; SPURIOUS;
Start;9000.000000;Hz
Stop;7000000000.000000;Hz
x-Axis;LIN;
Sweep Count;1;
The data section for the measurement values starts with the key word "TRACE <n>:
", where <n> includes the number of the used trace. Next comes the peak list with
several columns also separated by ';'.
Table 4-2
File contents
TRACE 1:
Trace Mode;CLR/WRITE;
x-Unit;Hz;
y-Unit;dBm;
Margin;6.000000:s
Values;8;
1;1548807257.5999999000;-
65.602280;-5.602280
1;1587207214.4000001000;-
65.327530;-5.327530
1;2112006624.0000000000;-
4.388008;55.611992
Spreadsheet programs such as MS Excel can read this format. Use ';' as the delim-
iter for the table cells.
Different language versions of evaluation programs may require different handling
of the decimal point. Using the DECIM SEP softkey, you can thus choose between
the delimiters '.' (decimal point) and ',' (comma).
Remote command:
4.222
Example: ASCII export table - file header
Explanation
Model
Firmware version
Storage date of data set
Operating mode of the device
Start/stop of the span
Unit: Hz
Linear (LIN) or logarithmic (LOG) (future feature)
scaling of the x-axis
Selected number of sweeps
Example: ASCII export table - file data section
Explanation
Selected trace
Display mode of trace:
CLR/WRITE,AVERAGE,MAX HOLD,MIN HOLD,
VIEW, BLANK
Unit of x values:
Unit of y values:
Peak List margin
Number of measurement points
Measurement values:
<Trace>;<x value>; <y value>;<delta limit>
MMEM:STOR:SPUR,'F:\TEST.ASC'
Instrument Functions
Analyzer Mode
Operating Manual 1302.6163.12 - 03