R&S ESU Series Operating Manual page 474

Emi test receiver
Table of Contents

Advertisement

R&S ESU
CALCulate<1|2>:LIMit<1...8>:CONTrol:MODE RELative | ABSolute
This command selects the relative or absolute scaling for the x-axis of the selected limit line. The
definition is independent of the measurement window.
Example:
"CALC:LIM2:CONT:MODE REL"
'Defines the x-axis of limit line 2 as relatively scaled.
Characteristics: *RST value: ABSolute
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:CONTrol:SHIFt <numeric_value>
This command moves a limit line by the indicated value in x direction. In contrast to CALC:LIM:CONT:
OFFS, the line is shifted by modifying the individual x values and not by means of an additive offset.
The shift is independent of the measurement window.
In analyzer mode, the unit of values depends on the frequency or time domain of the x-axis, i.e. it is
HZ with CALC:LIM:CONT:DOM FREQ and S with CALC:LIM:CONT:DOM TIME.
In analyzer mode, the unit of values depends on the frequency or time domain of the x-axis, i.e. it is
HZ with CALC:LIM:CONT:DOM FREQ and S with CALC:LIM:CONT:DOM TIME.
Example:
"CALC:LIM2:CONT:SHIF 50KHZ"
'Shifts all reference values of limit line 2 by 50 kHz.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:LIMit<1...8>:CONTrol:SPACing LINear | LOGarithmic
This command selects linear or logarithmic interpolation for the calculation of limit lines from frequency
points.
Example:
"CALC:LIM:CONT:SPAC LIN"
Characteristics: *RST value: LIN
SCPI: device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.35