9Measuring Spurious Emissions - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
4.8.13.9
Measuring Spurious Emissions
All real amplifiers also generate unwanted RF products outside the assigned fre-
quency band. These spurious emissions are usually measured across a wide fre-
quency range, for example from 9 kHz to 7 GHz. The analyzer settings are specified
for each frequency range.
SPURIOUS
SPURIOUS ON | OFF
EMISSIONS
SWEEP LIST
LIST EVALUATION
IF SHIFT
PAGE UP / PAGE DOWN
START MEAS
STOP MEAS
Side menu
PEAK SEARCH
PEAKS PER RANGE
MARGIN
VIEW PEAK LIST
In the Spurious Emissions measurement mode, the R&S ESU performs measure-
ments in predefined frequency ranges with settings that can be specified individually
for each of these ranges.
For this purpose, the SWEEP TABLE settings or the current device settings are
used. Up to 20 subranges can be defined (they need not directly follow one another)
across which the R&S ESU sweeps in subsequent order. However, the measure-
ment ranges must not overlap. The measurement parameters can be selected inde-
pendently from each other in every subrange (SWEEP LIST menu, EDIT SWEEP
LIST).
4.216
!
EDIT SWEEP LIST
INS BEFORE RANGE
INS AFTER RANGE
DELETE RANGE
NEXT RANGES
PREVIOUS RANGES
ADJUST AXIS
START MEAS
STOP MEAS
!
!
SORT BY FREQUENCY
SORT BY DELTA LIM
ASCII FILE EXPORT
DECIM SEP
PAGE UP / PAGE DOWN
Instrument Functions
Analyzer Mode
Operating Manual 1302.6163.12 - 03

Hide quick links:

Advertisement

Table of Contents
loading