Factors - Selection of factors for the subrange
The permissible transducer factors for the marked subrange can be selected in a
selection box. Only factors matching with the unit of the set and fully covering the
selected subrange are permissible.
After each change of range limits, R&S ESU thus checks the factor list and, if
required, rebuilds it.
After reducing the start frequency or increasing the stop frequency of a range it may
happen that the factors defined for this range no longer fully cover the range. These
factors are deleted for this range when the transducer factor table is opened next
time.
A maximum of 4 transducer factors can be switched on at the same time in each
subrange. If none of them is switched on, 0 dB is assumed as a factor for the whole
subrange.
Remote command:
SAVE TRD SET
The SAVE TRD SET softkey saves the changed table in a file on the internal hard
disk. If a transducer name with the same name already exists, a corresponding
query is performed beforehand:
If the saved set is switched on, the new values will be used immediately.
Remote command:
4.9.3.5
Programming the Interface Configuration and Time Setup
The GENERAL SETUP softkey opens a submenu in which the general instrument
parameters can be set up. In addition to the configuration of the digital interfaces
(IECBUS, COM), the date and time may be entered.
The current settings are displayed in tabular form on the display screen where they
may be edited.
GENERAL
GPIB
SETUP
USER PORT
COM INTERFACE
4.278
--
-- (executed automatically)
!
GPIB ADDRESS
ID STRING FACTORY
ID STRING USER
GPIB LANGUAGE
IF GAIN NORM/PULS
SWEEP REP ON/OFF
COUPLING FSP/HP
!
USER PORT IN/OUT
PORT 0 0/1 to
PORT 7 0/1
Instrument Functions
Operating Manual 1302.6163.12 - 03