R&S ESU
Error code
-400
-410
-420
-430
-440
9.3 Device-Specific Messages
Table 9-1
Error code
1036
1052
1060
2022
2028
9.8
Error text in the case of queue poll
Error explanation
Query error
General error occurring when data are requested by a query.
Query INTERRUPTED
The query has been interrupted.
Example: After a query, the instrument receives new data before the response has
been sent completely.
Query UNTERMINATED
The query is incomplete.
Example: The instrument is addressed as a talker and receives incomplete data.
Query DEADLOCKED
The query cannot be processed.
Example: The input and output buffers are full, the instrument cannot continue
operation.
Query UNTERMINATED after indefinite response
A query is in the same command line after a query which requests an indefinite
response.
Device-specific messages
Error text in the case of queue poll
Error explanation
MS: The correction table based amplifier gain exceeds the amplifier range for
CALAMP1 and CALAMP2 on IF board
This message is output when the setting range of the calibration amplifier is not
sufficient for the required correction. The error occurs only with modules which are not
correctly adjusted or defective.
Frontend LO is Unlocked
This message is output when the phase regulation of the local oscillator fails in the RF
frontend.
Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is output when the gate signal length is not sufficient for the pull-in delay
with a predefined gate delay.
OPTIONS.INI invalid
This message is output when an error has been recognized in the file OPTIONS.INI
which contains the clearing codes for retrofitable firmware applications. If this file is not
correctly recognized, all firmware applications are blocked for this instrument.
Hardcopy not possible during measurement sequence
This message is output when a printout is started during scan sequences that cannot
be interrupted. Such sequences are for example:
• Recording the system error correction data (calibration)
• Instrument selftest
In such cases a synchronization to the end of the scan sequence should be performed
prior to starting the printout.
Error Messages
Device-Specific Messages
Operating Manual 1302.6163.12 - 03