Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:FREQuency:SWEep[:STATe] ON | OFF
This command activates or deactivates the frequency sweep for generator 1 or 2 in the selected
window.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT1:FREQ:SWE ON"
'Activates the frequency sweep for ext. generator 1.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:POWer[:LEVel] <numeric_value>
This command sets the output power of the selected generator 1 or 2 in the selected window.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT:POW –30dBm"
'Sets the generator level to –30 dBm
Characteristics: *RST value: -20 dBm
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:ROSCillator[:SOURce] INTernal | EXTernal
This command switches between external and internal reference oscillator for the frequency
processing of external generator 1 and 2.
The command always works on both generators. Therefore, the numeric suffix in EXTernal<1|2> is not
significant.
This command is only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example:
"SOUR:EXT:ROSC EXT"
'switches to external reference oscillator
Characteristics: *RST value: INT
SCPI: device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SOURce Subsystem