4.9.3.2
Control of V-Networks (LISNs)
LISN
The LISN softkey opens the submenu including the settings required to control V-
networks (LISNs).
ESH2-Z5 / ESH3-Z5 / ENV 4200 / ENV 216 / OFF
PHASE N / PHASE L1 / PHASE L2 / PHASE L3
PE GROUNDED / PE FLOATING
ESH2-Z5 /
The ESH2-Z5, ESH3-Z5, ENV 4200, ENV 216 and OFF softkeys select the V-net-
ESH3-Z5 /
work to be controlled via the user port. They are toggle softkeys, and only one of
ENV 4200 /
them can be activated at a time.
ENV 216 / OFF
ESH2-Z5
ESH3-Z5
ENV 4200
ENV 216
OFF
Remote command:
PHASE N /
The PHASE N, PHASE L1, PHASE L2, and PHASE L3 softkeys select the phase of
PHASE L1 /
the V-network on which the RFI voltage is to be measured.
PHASE L2 /
PHASE N
PHASE L3
PHASE L1
PHASE L2
PHASE L3
Remote command:
150 KHZ
The 150 KHZ HIGHPASS softkey switches an 150 kHz highpass on to protect the
HIGHPASS
receiver input from high signal levels below 150 kHz (only for ENV 216).
Remote command:
PE GROUNDED
The PE GROUNDED and PE FLOATING softkeys switch the protective earth con-
/
ductor chokes on or off.
PE FLOATING
PE GROUNDED
PE FLOATING
Remote command:
Operating Manual 1302.6163.12 - 03
four-line V-network is controlled.
two-line V-network is controlled.
four-line V-network is controlled.
two-line V-network is controlled.
remote control is deactivated.
INP:LISN TWOP | FOUR | ENV4200 | ENV216 |
OFF
RFI on phase N is measured.
RFI on phase L1 is measured.
RFI on phase L2 is measured
(only for ESH2-Z5/ENV 4200).
RFI on phase L3 is measured
(only for ESH2-Z5/ENV 4200).
INP:LISN:PHAS L1|L2|L3|N
INP:LISN:FILT:HPAS ON | OFF
Protective earth conductor choke is switched off.
Protective earth conductor choke is switched on.
INP:LISN:PEAR GRO|FLO