6.5.6
CALCulate:PEAKsearch I PSEarch Subsystem
The CALCulate:PEAKsearch subsystem contains commands for verifying scan results in the test
receiver mode.
CALCulate<1|2>:PEAKsearch|PSEarch:AUTO
By using this command, the peak list in the spurious measurement is calculated automatically after a
measurement. For each range, exactly one peak value is calculated.
The SENSe suffix is unused.
Example:
"CALC:PEAK:AUTO ON"
'Switches the automatic peak search on.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
CALCulate<1|2>:PEAKsearch|PSEarch:ADD
This command adds a frequency value to the peaklist in receiver mode (RECEIVER – FINAL MEAS –
EDIT PEAK LIST). Only frequencies in the currently displayed frequency range of the receiver scan
are allowed.
Example:
"CALC:PEAK:ADD 93MHz"
'The frequency 93 MHz as added to the peak list.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R
CALCulate<1|2>:PEAKsearch|PSEarch:CLEar[:IMMediate]
The peaklist in receiver mode (RECEIVER – FINAL MEAS – EDIT PEAK LIST) is cleared.
Example:
"CALC:PEAK:CLE"
'The peak list is cleared.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R
CALCulate<1|2>:PEAKsearch|PSEarch[:IMMediate]
This command activates the generation of a peak list.
The numeric suffix in CALCULATE<1|2> is not significant.
Example:
"CALC:PEAK"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A, R
This command is an event and therefore has no *RST value and no query.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
ON | OFF
<nummeric value>
CALCulate Subsystem
6.89