R&S ESU
MMEMory:SELect[:ITEM]:NONE
This command deletes all data subsets from the list of device settings to be stored/loaded.
Example:
"MMEM:SEL:NONE"
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value.
MMEMory:SELect[:ITEM]:SCData ON | OFF
This command adds the tracking generator calibration data to the list of device settings to be stored/
loaded.
Example:
"MMEM:SEL:SCD ON"
'Inserts the tracking generator correction data in the list of data subsets
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
This command is only available in conjunction with the tracking generator option B9 or external
generator control option B10.
MMEMory:SELect[:ITEM]:TRACe[:ACTive] ON | OFF
This command adds the active traces to the list of data subsets of a save/recall device setting. Active
traces are all traces whose state is not blank.
Example:
"MMEM:SEL:TRAC ON"
Characteristics: *RST value: OFF, i.e. no traces will be stored
SCPI: device-specific
Mode:
R, A
MMEMory:SELect[:ITEM]:TRANsducer:ALL ON | OFF
This command includes all transducer factors and sets in the list of data subsets of a device setting to
be stored/loaded.
Example:
"MMEM:SEL:TRAN:ALL ON"
Characteristics: *RST value: ON
SCPI: device-specific
Mode:
R, A
MMEMory:STORe:FINal <file_name>
This command stores the final measurement data in an ASCII file.
The file name includes indication of the path and may also include the drive. The path name
corresponds to the DOS conventions.
Parameter:
<file_name> := DOS file name
Example:
":MMEM:STOR:FIN 'A:\TEST.ASC'"
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
MMEMory Subsystem
6.147