R&S ESU
Table 4-6
Accuracy
High
Medium
-
At a reference level (REF LEVEL) of -10 dBm and at a tracking generator output
level of the same value the R&S ESU operates without overrange reserve, i.e. the
R&S ESU is in danger of being overloaded if a signal is applied whose amplitude
is higher than the reference line. In this case, either the message OVLD for over-
load is displayed in the status line or the display range is exceeded (clipping of the
trace at the upper diagram border = Overrange).
Overloading can be avoided as follows:
Reducing the output level of the tracking generator (SOURCE POWER,
NETWORK menu)
Increasing the reference level (REF LEVEL, AMPT menu)
4.328
Measurement accuracy levels
Enhancement
Reason/Limitation
label
NOR
No difference between reference setting and
measurement
APX
Change of the following settings:
(approximation)
• coupling (RBW, VBW, SWT)
• reference level, RF attenuation
• start or stop frequency
• output level of tracking generator
• frequency offset of tracking generator
• detector (max. peak, min. peak, sample, etc.)
Change of frequency:
• max. 625 points within the set sweep limits
(corresponds to a doubling of the span)
Aborted
• 625 and more extrapolated points within the current
normalization
sweep limits (in case of span doubling)
Instrument Functions
Tracking Generator – Option R&S FSU-B9
Operating Manual 1302.6163.12 - 03