R&S ESU Series Operating Manual page 576

Emi test receiver
Table of Contents

Advertisement

R&S ESU
INSTrument:COUPle:PROTection
This command couples the RF protection attenuation of 10 dB for the analyzer and receiver modes
with one another.
To prevent damage to the RF input, activate this coupling.
Example:
"INST:COUP:PROT ALL"
Characteristics: *RST value:ALL
SCPI:device-specific
Mode:
R, A
INSTrument:COUPle:SPAN
This command couples the start and stop frequencies of sweep and scan for the analyzer and receiver
modes with one another.
Example:
"INST:COUP:SPAN ALL"
Characteristics: *RST value:NONE
SCPI:device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
ALL | NONE
ALL | NONE
Remote Control – Description of Commands