Mode:
R, A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:MODE:HCONtinuous ON | OFF
This command specifies whether or not the traces with peak or minimum value detection are reset after
specific parameter changes.
Usually the measurement must be restarted after a parameter change, before an evaluation of the
measurement results is performed (e.g. with a marker). In cases in which a change causes a
compulsory new measurement, the trace is automatically reset in order to prevent erroneous
measurements of previous measurement results (e.g. when the span changes). For applications in
which this behavior is not desired, this mechanism can be switched off.
Parameter:
OFF: The traces are reset after specific parameter changes.
ON: The reset mechanism is switched off.
Example:
"DISP:WIND1:TRAC3:MODE:HCON ON"
'The reset mechanism is switched off for measurement window 1.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>[:STATe] ON | OFF
This command switches on or off the display of the corresponding trace in the selected measurement
window.
Example:
"DISP:WIND1:TRAC3 ON"
Characteristics: *RST value: ON for TRACe1, OFF for TRACe2 and 3
SCPI: conform
Mode:
R, A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:SYMBol CROSs | OFF
This command switches on or off the indication of the peak list or the final measurement result on the
trace.
Parameter:
CROSs: The single value is indicated as x.
OFF: No indication.
Example:
"DISP:TRAC:SYMB CROS"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R
DISPlay:BARGraph:LEVel:LOWer?
This command queries the minimum level of the bar graph.
Example:
":DISP:BARG:LEV:LOW?"
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R