CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer[:STATe] OFF
This command switches off the power measurement in the selected measurement window.
Example:
"CALC:MARK:FUNC:POW OFF"
'Switches off the power measurement in screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-F
This command is an event and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:MODE WRITe | MAXHold
This command selects the Clear Write or Maxhold for Channel Power values.
Example:
"CALC:MARK:FUNC:POW:MODE MAXH"
'Maxhold for Channel Power values
Characteristics: *RST value: WRITe
SCPI: device-specific
Mode:
A-F
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:PRESet NADC | TETRA | PDC | PHS | CDPD |
FWCDma | RWCDma | F8CDma | R8CDma | F19Cdma | R19Cdma | FW3Gppcdma | RW3Gppcdma
| D2CDma | S2CDma | M2CDma | FIS95A | RIS95A | FIS95C0 | RIS95C0 | FJ008 | RJ008 | FIS95C1
| RIS95C1 | TCDMa | NONE | AWLan | BWLan | WIMax | WIBro | EUTRa | REUTra | <string>'
This command selects the power measurement setting for a standard in the indicated measurement
window and previously switches on the corresponding measurement, if required. The function is
independent of the marker selection, i.e. the numeric suffix <1...4> of MARKer is irrelevant.
The configuration for a standard comprises of the parameters weighting filter, channel bandwidth and
spacing, resolution and video bandwidth, as well as detector and sweep time
Meaning of the CDMA standard abbreviations:
FIS95A, F8CDma
CDMA IS95A forward
RIS95A, R8CDma
CDMA IS95A reverse
FJ008, F19CDma
CDMA J-STD008 forward
RJ008, R19CDma
CDMA J-STD008 reverse
FIS95C0
CDMA IS95C Class 0 forward
RIS95C0
CDMA IS95C Class 0 reverse
FIS95C1
CDMA IS95C Class 1 forward
FIS95R1
CDMA IS95C Class 1 reverse
FWCDma
W-CDMA 4.096 MHz forward
RWCDma
W-CDMA 4.096 MHz reverse
FW3Gppcdma
W-CDMA 3.84 MHz forward
RW3Gppcdma
W-CDMA 3.84 MHz reverse
D2CDma
CDMA2000 direct sequence
S2CDma
CDMA2000 MC1 multi carrier with 1 carrier
6.70
Remote Control – Description of Commands