R&S ESU
Example:
"CALC2:DELT:FUNC:FIX ON"
'Switches on the measurement with fixed reference value for all delta markers in
screen B.
"CALC2:DELT:FUNC:FIX:RPO:X 128 MHZ"
'Sets the reference frequency in screen B to 128 MHz.
"CALC2:DELT:FUNC:FIX:RPO:Y 30 DBM"
'Sets the reference level in screen B to +30 dBm
Characteristics: *RST value: OFF
SCPI: device-specific.
Mode:
A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
<numeric_value>
This command sets the reference point level for all delta markers in the selected measurement window
for a measurement with fixed reference point (CALC:DELT:FUNC:FIX:STAT ON) to the peak of the
selected trace.
For phase noise measurements (CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON), the
command defines a new reference point level for delta marker 2 in the selected measurement window.
Example:
"CALC:DELT:FUNC:FIX:RPO:MAX"
'Sets the reference point level for the delta marker in screen A to the peak of the
trace.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y <numeric_value>
This command defines a new reference point level for all delta markers in the selected measurement
window for a measurement with fixed reference point. (CALCulate:DELTamarker:FUNCtion:
FIXed:STATe ON).
For phase noise measurements (CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON), the
command defines a new reference point level for delta marker 2 in the selected measurement window.
Example:
"CALC:DELT:FUNC:FIX:RPO:Y -10dBm"
'Sets the reference point level for delta markers in screen A to -10 dBm.
Characteristics: *RST value: - (FUNction:FIXed[:STATe] is set to OFF)
SCPI: device-specific
Mode:
A
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y:OFFSet <numeric_value>
This command defines an additional level offset for the measurement with fixed reference value
(CALCulate:DELTamarker:FUNCtion:FIXed:STATe ON). For this measurement, the offset is
included in the display of all delta markers of the selected measurement window.
For phase noise measurements (CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON), the
command defines an additional level offset which is included in the display of delta marker 2 in the
selected measurement window.
6.18
Remote Control – Description of Commands
CALCulate Subsystem
Operating Manual 1302.6163.12 - 03