R&S ESU
The COMMENT SCREEN A / B softkeys allow text to be added to the printout (date
and time are inserted automatically).
The REPORT softkey is used to configure and print test reports. In contrast to the
hardcopy which is restricted to the contents of the screen, the test report consists of
more different items, such as a diagram, a header, a scan table and lists with mea-
surement results. For detailed information refer to section
Report" on page
Use the INSTALL PRINTER softkey to install additional printer drivers.
PRINT
The PRINT SCREEN softkey starts the output of test results.
SCREEN
All the diagrams, traces, markers, marker lists, limit lines etc. are printed out as long
as they are displayed on the screen. All the softkeys, tables and open data entry
fields are not printed out. Moreover, comments, title, date, and time are output at the
bottom margin of the printout.
Remote command:
PRINT TRACE
The PRINT TRACE softkey starts the output of all curves visible on the display
screen without auxiliary information. Specifically, no markers or display lines are
printed.
Remote command:
PRINT TABLE
The PRINT TABLE softkey starts the output of all tables and info lists visible on the
display screen without the measurement diagrams and other information lying
behind.
Remote command:
DEVICE
The DEVICE SETUP softkey opens the dialog where the file format and the printer
SETUP
can be selected. For details on selecting and configuring printers, and printing out
the measurement results refer to the Quick Start Guide chapter 1, "Preparing for
Use" and chapter "Printing Out Measurement Results".
Remote command:
DEVICE 1 / 2
The R&S ESU is able to manage two hardcopy settings independently of each
other. They are selected via the DEVICE 1 / 2 softkey, which displays also the asso-
ciated setting if the DEVICE SETUP dialog is open. For details on selecting and con-
figuring printers, and printing out the measurement results refer to the Quick Start
Guide chapter 1, "Preparing for Use" and chapter "Printing Out the Measurement
Results".
Remote command:
Operating Manual 1302.6163.12 - 03
4.313.
HCOP:ITEM:ALL
HCOP:IMM
HCOP:ITEM:WIND:TRAC:STAT ON
HCOP:IMM
HCOP:ITEM:WIND:TABL:STAT ON
HCOP:IMM
HCOP:DEV:LANG GDI;
SYST:COMM:PRIN:ENUM:FIRS?;
SYST:COMM:PRIN:ENUM:NEXT?;
SYST:COMM:PRIN:SEL <Printer>;
HCOP:PAGE:ORI PORT;
HCOP:DEST "SYST:COMM:PRIN";
HCOP:DEST "SYST:COMM:MMEM"
--