Characteristics: *RST value: -
SCPI: conform
Mode:
A
CALCulate<1|2>:LIMit<1...8>:CLEar[:IMMediate]
This command deletes the result of the current limit check for all limit lines.
Up to 8 limit lines can be defined at the same time. The numeric suffixes <1|2> of CALCulate indicate
the measurement window.
This command is an event and therefore has no *RST value.
Example:
"CALC:LIM:CLE"
'Deletes the result of the limit check in screen A.
Characteristics: *RST value: -
SCPI: conform
Mode:
A
CALCulate<1|2>:LIMit<1...8>:COMMent <string>
This command defines a comment for the limit line selected (max. 40 characters).
Up to 8 limit lines can be defined at the same time. The comment is independent from the
measurement window.
Example:
"CALC:LIM5:COMM 'Upper limit for spectrum'"
'Defines the comment for limit line 5.
Characteristics: *RST value: blank comment
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:COPY 1 to 8 | <name>
This command copies one limit line onto another one.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Parameter:
1 to n ::= number of the new limit line or
<name> ::= name of the new limit line given as a string
Example:
"CALC:LIM1:COPY 2"
'Copies limit line 1 to line 2.
"CALC:LIM1:COPY 'FM2'"
'Copies limit line 1 to a new line named 'FM2'.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.25