R&S ESU Series Operating Manual page 579

Emi test receiver
Table of Contents

Advertisement

R&S ESU
MMEMory:CLEar:ALL
This command deletes all device settings in the current directory. The current directory can be selected
with MMEM:CDIR. The default directory is D:.
Example:
"MMEM:CLE:ALL"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
MMEMory:COMMent <string>
This command defines a comment (max. 60 characters) for a device setting to be stored.
Example:
"MMEM:COMM 'Setup for FM measurement'"
Characteristics: *RST value: blank comment
SCPI: device-specific
Mode:
R, A
MMEMory:COPY <file_source>,<file_destination>
This command copies the files indicated in <file_source> to the destination directory indicated with
<file_destination> or to the destination file indicated by <file_destination> when <file_source> is just a
file.
The indication of the file name may include the path and the drive name. The file names and path
information must be in accordance with the DOS conventions.
Parameter:
<file_source>,<file_destination> ::= <file_name>
<file_name> ::= DOS file name
Example:
"MMEM:COPY 'D:\USER\DATA\SETUP.CFG','F:'"
Characteristics: *RST value: -
SCPI: conform
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
MMEMory:DATA <file_name>[,<block data>]
This command writes the block data contained in <block> into the file characterized by <file_name>.
The GPIB delimiter must be set to EOI to obtain error-free data transfer.
The associated query command reads the indicated file from the mass memory and transfers it to the
control computer via the GPIB. It should be noted that the buffer memory of the control computer
should be large enough to store the file. The setting of the GPIB delimiter is irrelevant in this case.
The command is useful for reading stored device settings or trace data from the instrument or for
transferring them to the instrument.
MMEMory:DATA <file_name>,<block data>
Data transfer from control computer to instrument.
MMEMory:DATA? <file_name>
Data transfer from instrument to control computer.
6.140
Remote Control – Description of Commands
MMEMory Subsystem
Operating Manual 1302.6163.12 - 03

Hide quick links:

Advertisement

Table of Contents
loading