The function is identical to softkey RANGE LOG MANUAL in AMPT menu.
Remote command:
Y-UNIT %/ABS
The softkey Y-UNIT %/ABS defines the scaling type on the y-axis. The default case
are the absolute probability. This can be changed to percent values. The softkeys Y-
AXIS MIN and Y-AXIS MAX are using values based on the Y-UNIT setting.
Remote command:
The 0.01%, 0.1%, 1% and 10% value from the CCDF measurement are shown in
the bottom screen half. Those values can also queried via remote:
Remote command:
Y-AXIS MAX
The Y-AXIS MAX VALUE softkey defines the upper limit of the displayed probability
VALUE
range.
Values on the y-axis are normalized which means that the maximum value is 1.0. As
the y-axis scaling has a logarithmic axis the distance between max and min value
must be at least one decade.
Remote command:
Y-AXIS MIN
The Y-AXIS MIN VALUE softkey defines the lower limit of the displayed probability
VALUE
range.
As the y-axis scaling has a logarithmic axis the distance between max and min value
must be at least one decade. Valid values are in the range 0 < value < 1.
Remote command:
ADJUST
The ADJUST SETTINGS softkey optimizes the level settings of the R&S ESU
SETTINGS
according to the measured peak power in order to gain maximum sensitivity of the
instrument.
The level range is adjusted according to the measured difference between peak and
minimum power for APD measurement and peak and mean power for CCDF mea-
surement in order to obtain maximum power resolution.
Additionally the probability scale is adapted to the selected number of samples.
Remote command:
DEFAULT
The DEFAULT SETTINGS softkey resets the x- and y-axis scalings to their PRESET
SETTINGS
values.
x-axis ref level: -20 dBm
x-axis range APD: 100 dB
x-axis range CCDF: 20 dB
y-axis upper limit: 1.0
y-axis lower limit: 1E-6
Remote command:
4.204
CALC:STAT:SCAL:X:RANG <value>
CALC:STAT:SCAL:Y:UNIT PCT | ABS
CALC:STAT:CCDF:X? P0_01 | P0_1 |
P1 | P10
CALC:STAT:SCAL:Y:UPP <value>
CALC:STAT:SCAL:Y:LOW <value>
CALC:STAT:SCAL:AUTO ONCE
CALC:STAT:PRES