R&S ESU Series Operating Manual page 515

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak:AVERage:RESult?
This command is used to query the result of the measurement of the averaged positive peak value in
the selected measurement window. The query is only possible if averaging has been activated
previously using CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:AVERage.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> in MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:AVER ON"
'Switches on the calculation of average in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:AVER:RES?"
'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
6.76
Remote Control – Description of Commands
CALCulate Subsystem
Operating Manual 1302.6163.12 - 03

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents