R&S ESU Series Operating Manual page 629

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]LIST:RANGe<1...20>[:FREQuency]:STOP <numeric_value>
This command sets the stop frequency of a range in the spurious measurement.
Example:
"LIST:RANG2:STOP 2GHZ"
'Stop frequency of range 2 to 2 GHz
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:INPut:ATTenuation <numeric_value>
This command defines the RF attenuation of a range in the spurious measurement.
Example:
"LIST:RANG2:INP:ATT 30db"
'RF attenuation of range 2 to 30 dB
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:INPut:ATTenuation:AUTO ON | OFF
This command switches the autoranging of a range in the spurious measurement on or off.
Example:
"LIST:RANG2:INP:ATT:AUTO ON"
'Activates autorange for range 2
Characteristics: *RST value: ON
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:INPut:GAIN:STATe ON | OFF
This command switches the preamplifier of a range in the spurious measurement on or off.
Example:
"LIST:RANG2:INP:GAIN:STAT ON"
'Activates the preamplifier for range 2
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:LIMit -200dB to +200dB
This command defines the limit lines for the spurious emission measurement
details on this measurement see page 220). For each of the 20 ranges, you can specify a limit line via
the numeric suffix at RANGe<1...20>. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"LIST:RANG5:LIM -40"
'Limit line in range 5 is –40 dB.
Characteristics: *RST value: -13dB
SCPI: device-specific
Mode:
A
6.190