R&S ESU Series Operating Manual page 282

Emi test receiver
Table of Contents

Advertisement

R&S ESU
2nd press: The line is switched off (softkey with gray background).
Initial situation: The line is on (softkey with green background)
1st press: The data input function is activated (softkey with red background). The
position of the display line can be selected by means of the rotary knob, the step
keys or a numerical entry in the appropriate field. The data input function is disabled
if another function is activated. The line, however, remains switched on (softkey with
green background).
2nd press: The line is switched off (softkey with gray background).
DISPLAY
DISPLAY LINE 1 /
LINES
DISPLAY LINE 2
FREQUENCY LINE 1/
FREQUENCY LINE 2
TIME LINE 1 /
TIME LINE 2
DISPLAY LINE 1
The DISPLAY LINE 1/2 softkeys enable or disable the level lines and allow the user
/
to enter the position of the lines.
DISPLAY LINE 2
The level lines mark the selected level in the measurement window.
These softkeys are only available in the time domain (span = 0).
Remote command:
FREQUENCY
The FREQUENCY LINE 1/2 softkeys enable or disable the frequency lines 1/2 and
LINE 1/
allow the user to enter the position of the lines.
FREQUENCY
The frequency lines mark the selected frequencies in the measurement window.
LINE 2
These softkeys are only available in the frequency domain (span > 0).
Remote command:
TIME LINE 1 /
The TIME LINE 1/2 softkeys enable or disable the time lines 1/2 and allow the user
TIME LINE 2
to enter the position of the lines.
The time lines mark the selected times or define search ranges (see section
Functions – MKR FCTN Key" on page
The two softkeys cannot be used in the frequency domain (span > 0).
Operating Manual 1302.6163.12 - 03
CALC:DLIN:STAT ON
CALC:DLIN -20dBm
CALC:FLIN:STAT ON
CALC:FLIN 120MHz
4.152).
Instrument Functions
Basic Settings
"Marker
4.251

Hide quick links:

Advertisement

Table of Contents
loading