R&S ESU
The number of sweeps set in the TRACE menu is the same as that in the SWEEP
menu.
If SINGLE SWEEP is selected, the measurement stops after the selected number
of sweeps has been performed.
Remote command:
SWEEP
The SWEEP POINTS softkey selects the number of measurement samples
POINTS
acquired during a sweep.
The following numbers of points per sweep are available: 155, 201, 301, 313, 401,
501, 601, 625 (default), 701, ..., 30001
An increment of 100 is possible for number of points ≥ 201.
Additionally the numbers 1251 and 1999 are available.
The autopeak detector will be disabled while the number of points per sweep is set
to another value than the default.
Remote command:
SGL SWEEP
The SGL SWEEP DISP OFF softkey deactivates the display while a single sweep is
DISP OFF
being performed. Once the sweep has been completed, the trace is shown.
Remote command:
4.8.6
Triggering the Sweep – TRIG Key
The TRIG key opens a menu for selection of the various trigger sources, trigger
polarity and external gate function. The active trigger mode is indicated by highlight-
ing the corresponding softkey.
For video trigger, a trigger threshold can be entered, which is represented in the dia-
gram as a horizontal line.
To indicate that a trigger mode other than FREE RUN has been set, the enhance-
ment label TRG is displayed on the screen. If two windows are displayed, TRG
appears next to the appropriate window.
TRIG
FREE RUN
VIDEO
EXTERN
IF POWER
TRIGGER OFFSET
POLARITY POS | NEG
GATED TRIGGER
GATE SETTINGS
4.122
SWE:COUN 64
SWE:POIN 501
INIT:DISP OFF
!
GATE MODE LEVEL | EDGE
POLARITY POS | NEG
Instrument Functions
Analyzer Mode
Operating Manual 1302.6163.12 - 03