Operation On A Discrete Frequency - Freq Key - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
4.5.1
Operation on a Discrete Frequency – FREQ Key
FREQ
The FREQ key opens the FREQUENCY menu for setting the receive frequency in
manual operation and the frequency axis for scan display.
RECEIVER FREQUENCY
STEPSIZE
START
STOP
RUN PRE-SCAN+FINAL
RUN SCAN
RECEIVER
The RECEIVER FREQUENCY softkey activates the entry field of the receiver fre-
FREQUENCY
quency.
The tuning frequency has to be set to at least twice the IF bandwidth.
When the tuning frequency is lower than twice the IF bandwidth, the IF bandwidth is
automatically reduced so that this condition is met again.
If the frequency is increased again, the original IF bandwidth is restored (memory
function). The memory is cleared when the IF bandwidth is manually changed.
Range: 20 Hz ≤ f
This softkey is also available in the RECEIVER / MEAS menu.
Remote command:
STEPSIZE
The STEPSIZE softkey opens a submenu for setting the step size of the receive fre-
quency. The step size can be coupled to the set frequency or be manually set to a
fixed value. The softkeys of the submenu are mutually exclusive selection switches.
Only one switch can be activated at any one time.
AUTO COARSE
If the AUTO COARSE softkey is activated, the receive frequency is set in coarse
steps. The 4th digit of the selected frequency is varied.
Remote command:
AUTO FINE
If the AUTO FINE softkey is activated, the receive frequency is set in fine steps. The
7th digit of the selected frequency is varied.
Remote command:
STEPSIZE
The STEPSIZE MANUAL softkey activates the entry window for the input of a fixed
MANUAL
step size.
Remote command:
4.14
!
AUTO COARSE
AUTO FINE
STEPSIZE MANUAL
STEPSIZE = FREQ
≤ f
rec
max
FREQ:CENT 300 MHz
--
--
FREQ:CENT:STEP 50 kHz