R&S ESU Series Operating Manual page 267

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Sweep Time Mode
AUTO, MANUAL
Remote command:
Sweep Time
Shows the current sweep time. The R&S ESU automatically calculates the sweep
time if AUTO mode is selected and displays the corresponding value in this field.
If the field is edited, the sweep time mode is automatically set to MANUAL.
Remote command:
REF Level
Reference level in dBm. The uppermost grid line in the measurement diagram
represents the reference level, corrected by the associated transducer factor.
The unit depends on the unit selected via the UNIT softkey.
Remote command:
RF Attenuator Mode
AUTO, MANUAL
Remote command:
RF Attenuator
Shows the current RF attenuation. The R&S ESU automatically calculates the RF
attenuation if AUTO mode is selected and displays the corresponding value in this
field. If the field is edited, the RF Attenuator mode is automatically set to MANUAL.
Remote command:
PRE AMP
Remote command:
Transd. Factor
Shows the name of the transducer factor that is used for the measurement, or
NONE if no transducer factor is selected.
To select a transducer factor highlight this field with the rotary knob and press the
ENTER key to open the transducer selection list. Select the name of the
transducer factor you need with the rotary knob or NONE if you have to work
without a transducer factor.
Remote command:
4.236
ESP:RANG1:SWE:TIME:AUTO ON
ESP:RANG1:SWE:TIME 1
ESP:RANG1:RLEV 0
ESP:RANG1:INP:ATT:AUTO ON
ESP:RANG1:INP:ATT 10
ESP:RANG1:INP:GAIN:STAT ON
ESP:RANG1:TRAN <string>
Operating Manual 1302.6163.12 - 03
Instrument Functions
Analyzer Mode

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents